Functional Data
Principle High performance table-top flatbed scanner with
contactless plate transport
Eraser Inline
Resolution Standard resolution (SR) 70 µm
High resolution (HR) 35 µm
Maximum basic Standard resolution (SR) 80 µm (6,25 LP / mm)
Spatial resolution High resolution (HR) 40 µm (12,5 LP / mm)
Scan width 35 cm (14 inch)
Throughput
(10 x 40 cm
4.5 x 17”)
Standard resolution (SR) 90 plates/hour
High resolution (HR) 28 plates/hour
Time to image
(in Rhythm RT)
Standard resolution (SR) 40 sec
High resolution (HR) 147 sec
LUT
(look up table)
Linear
(native Square root)
Bit depth 16 Bit
Dimensions Scanner 56 x 56 x 47 cm
(22” x 22” x 19”)
Scanner including I/O
table and light cover
128 x 56 x 47 cm
(52” x 22” x 19”)
Weight Scanner 45 kg (99 lbs)
Scanner including I/O
table and light cover
50 kg (110 lbs)
Interfaces Ethernet, RJ45
DC voltage, coded 8-pin, female
Certifications CE, UL (NRTLus), cUL (cNRTLus), C-Tick, Customs
Union Mark
Environmental Conditions
Operation Temp. allowed 15 °C to 35 °C
(59 °F to 95 °F)
Relative humidity 15% to 80%
(non condensed)
Magnetic field Compliant with
EN 61000-4-8, Level 2
Transport IEC721-3-2 (1997): class 2K2 and 2M3, with
following restrictions
Temperature -25 °C to +55 °C
(-13 °F to 131 °F)
Vibration 5 to 200 Hz
(vertical, longitudinal,
transversal axis)
Mechanical
conditions for
transport
In packaging IEC 721-3-2 (1997):
class 2M2
Shock
specifications
In packaging IEC TR 60721-4-5 (1997):
class 5M2
Electrical Data
Operating voltage Auto-ranging external power supply from
100 V to 240 V, DC Output 24V
Mains frequency 50/60 Hz
Mains fuse protection Europe min. 10 A, max. 16 A
USA & Japan min. 10 A, max. 15 A
Power
consumption
Standby
110 V - 240 V / 50-60 Hz
max. 22 W
During operation
110 V - 240 V / 50-60 Hz
max. 140 W
(absolute peak)
Application Compliance
ASME ASME Code Section V Article 2
ISO 17636-2 Class A / Class B
(in defined exposure
conditions)
Verified with X-ray,
Ir-192, Se-75, Co-60
EN14784-1 IPS, IPS-2: 1/80,
IPU: 1/40
Certified by BAM
EN2446-06 IPS, IPS-2: S/80,
IPU: S/40
Certified by BAM
Accessories
I/O Table with light cover Quick mountable, stainless steel, input/output
table set with 43 cm (17”) tray length and light
cover for input side
Long I/O table Input/output extension for long plates scanning
150 cm (59”)
Flight Case Robust Flight Case with shock-absorbers, wheel,
ruggedized handles and compartments for I/O
tables, laptop, accessories
Imaging Plates
IPC2 High speed plate Use: CRxVision can scan
any shape or size
imaging plate from 20
to 1500 mm (0.75 to 60”)
in length.
IPS High resolution
IPS2 High resolution
IPU Extremely high
resolution (X-ray)
Cassettes
Flexible cassettes PVC or vinyl envelopes different sizes
Hard cassettes
(for defined exposure conditions)
35 x 43 cm, 20 x 24 cm, 24 x 30 cm, 15 x 30 cm
14” x 17”, 8” x 10”, 10” x 12”, 6” x 12”
Technical Specifications of CRxVision
Imaging Plates
Four different types of imaging plates with a ferromagnetic
back layer have been developed specifically for the CRxVision.
This allows the imaging plates to be magnetically transported
through the scanner with no phosphor touch points. In addition,
this new design helps reduce backscatter which improves the
overall quality of the image as well as allows the imaging plate
the ability to return to a flat position after being constantly bent
around curved objects.
GE´s four imaging plate types are as follows:
• IPC2: Standard Resolution & High Speed - for general purpose
• IPS: High Resolution & Medium Speed - for inspection of welds
• IPS2: High Resolution & Medium Speed – for inspection of welds
Excellent for very low contrast, homogeneity type applications
and premium weld quality inspections.
• IPU: Extreme High Resolution & Slow Speed - for extremely high
resolution applications when very low micron range sensitivity
is required.
Plates are available in various size formats ranging from 70 mm
(2.76”) wide to 1500 mm (60”) in length.
Protective Cassettes
A range of flexible and hard cassettes are also available. Both
versions may be supplied with or without lead, depending on
the application.