Product Manual

Vscan maintenance
6-4 Vscan – User Manual
GM092207 02
Cleaning and disinfection
Cleaning
Cleaning the control unit, the display and the Docking station
Make sure the Docking station is disconnected from the AC
adapter before cleaning.
1. Switch off the Vscan from the Shutdown menu (see
page 3-2).
2. Moisten a soft, non-abrasive cloth with a mild, general
purpose, non-abrasive soap and water solution.
3. Wipe the Vscan and the Docking station.
4. Wipe dry with a soft towel.
NOTE: DO NOT scratch or press on any part of the Vscan with any
sharp objects, such as pencils or pens, as this may result in
damage of the Vscan.
Cleaning the probe
1. Remove the coupling gel by wiping the probe lens with a
soft cloth.
2. Wipe the probe and cable with a soft cloth moisten in a
warm soap and water solution (<80
o
F/27
o
C).
3. Wipe the probe and cable with a soft cloth moisten in clean
water (<80
o
F/27
o
C) until all soap is removed.
4. Wipe dry with a soft towel.
CAUTION
Do not spray any liquid directly onto the Vscan or the Docking
station.