Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
System Setup
3-40 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Wired Ethernet from Vivid E80/E90/E95 to a workstation
Either of these situations may apply:
• Direct Cable Connection from Vivid E80/E90/E95 to a
workstation via a Crossover Cable.
You will only need a Crossover Cable for the connection.
a. Connect one end of the Crossover Cable to the network
connector on the Vivid E80/E90/E95.
b. Connect the other end to the network connector on the
Workstation.
• Connection via a Peer-to-Peer network.
You will need a network hub and one network cable for each
Ultrasound system / workstation
• Connection via Hospital Network.
You will need one network cable to connect the Vivid E80/
E90/E95 to a wall outlet on the hospital’s network.
Connection from Vivid E80/E90/E95 to a DICOM Server on a network
You will need one network cable.
1. Connect one end of the cable to the Ethernet connector on
Vivid E80/E90/E95.
2. Connect the other end of the cable to the wall outlet.
NOTE: If a Peer-to-Peer network is connected to the hospital’s network,
you may connect the Vivid E80/E90/E95 to the Peer-to-Peer
network.