Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
System Setup
3-42 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Select TCP/IP Screen
Figure 3-9. TCP/IP Overview screen for Vivid E80/E90/E95
1. Press Config (F2) and log on as adm.
2. If not already selected, select Connectivity from the bottom row of “buttons” on the screen.
3. Select the TCP/IP TAB (named Tcpip).
The resulting screen gives you an overview of many of the network settings for Vivid E80/E90/E95.
1. My Computer:
• Computer Name:
For Vivid E80/E90/E95, this name is on the
form: VIVIDX-00NNNN, where “00NNNN” is a
number (NNNN is the Ultrasound system’s
serial number).
• AE Title:
VIVIDEx (where ‘x’ is the model number)
• Port No:
Default port number: 104
2. Server Config:
• Servers:
List of servers
• Buttons:
Use the buttons to Add, Modify, or Remove
servers.
3. Remote Path
Used for Save As, Export from Q-Analysis, and
exporting Error Logs with Alt-D.
4. Configurable Remote Path User:
Add Secondary Log-in Credential.
5. Save Settings:
Select Save Settings to archive any changes
you have done to the TCP/IP settings.
6. Network Settings:
Use Network Settings if you need to change
Vivid E80/E90/E95’s IP settings or turn DHCP
on or off.