Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
System Setup
3-48 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
To add a server in the Server Config list
Follow this procedure to add a server to the list:
1. Select Add.
Figure 3-16. Server Configuration
2. Add the server's name in the Server Name field.
3. Add the server’s IP address in the IP-Address field.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right of the
Check button indicates the result:
• A green check mark next to the Check button indicates
that the IP-Address was found on the network.
• A red mark indicates that the IP address was not found.
Troubleshooting tips:
• No network connection between the Ultrasound system
and the server.
• The wrong IP Address.
To modify the setup for a server in the list
NOTE: Both the pre-defined servers and new servers can be modified.
Follow this procedure to modify the setup for a server in the
list:
1. Highlight the server you want to modify.
2. Select Modify.
3. Update the information as needed.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right for the
Check button indicates the result:
• A green check mark next to the Check button indicates
that the IP-Address was found on the network.
• A red mark indicates that the IP address was not found.
Troubleshooting tips: