Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
General procedures
Vivid E80/E90/E95 – Service Manual 4-3
GC091052 Rev. 3
Connect AC (mains) Power to the Vivid E80/E90/E95
Connecting AC Power to the Vivid E80/E90/E95 ultrasound unit,
involves preliminary checks of the power cord, voltage level and
compliance with electrical safety requirements.
1. Ensure that the wall outlet is of appropriate type, and that
the Circuit Breaker is turned off.
2. Uncoil the power cable, allowing sufficient slack so that the
unit can be moved slightly.
3. Verify that the power cable is without any visible scratches
or any sign of damage.
4. Verify that the on-site mains voltage is within the limits
indicated on the rating label near the Circuit Breaker on the
rear of the unit.
5. Connect the Power Cable’s female plug to the Power Inlet at
the rear of the unit.
6. Lock the plug in position with the Retaining Clamp (ACC
Clamp).
7. Verify that the Mains Power Circuit Breaker is in OFF
position, if not, switch it OFF.
8. Connect the Power Cable’s other end (male plug) to a
hospital grade mains power outlet with the proper rated
voltage, and the unit is ready for Power ON/Boot Up.