Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Functional checks
Vivid E80/E90/E95 – Service Manual 4-11
GC091052 Rev. 3
Adjust the 2D mode controls
The following controls can be adjusted to optimize the 2D Mode
display:
• Swipe to page 2 on the Touch panel and press either Soft or
Sharp Auto Tissue setting.
• Soft: optimizes the radial and lateral uniformity and
brightness of the tissue continuously in real-time.
The mention “Soft” is displayed on the upper right
corner of the image area
• Sharp: further enhances the image display by
optimizing the grayscale curve.
The mention “Sharp” is displayed on the upper right
corner of the image area
The Auto Tissue setting (Soft or Sharp) can be turned on/off
by pressing Auto on the Control panel. The last used setting
is then applied.
The Auto Tissue settings are only available in live scanning
and cannot be turned off when the image is stored.
• If available, press Virtual Apex (probe dependent) to
improve near field imaging, allowing increased visibility up
to the width of the full probe aperture close to the surface.
• Use the Gain and TGC controls to optimize the overall
image.
Gain increases or decreases the amount of echo
information displayed. TGC compensates for depth-related
attenuation in the image.
WARNING
Always use the minimum power required to obtain acceptable
images in accordance with applicable guidelines and policies.