Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Functional checks
Vivid E80/E90/E95 – Service Manual 4-15
GC091052 Rev. 3
This unit has three types of M-Mode:
• Conventional M-Mode (MM): displays a distance/time plot of
a cursor line in the axial plane of the 2D-image.
• Anatomical M-Mode (AMM): displays a distance/time plot
from a cursor line, which is independent from the axial
plane. AMM is available in grayscale, color, TVI, Tissue
Tracking, Strain rate and Strain modes.
• Curved Anatomical M-Mode (CAMM): displays a distance/
time plot from a free-drawn cursor line. CAMM is available in
grayscale, color, TVI, Tissue Tracking, Strain rate and Strain
modes.
Conventional M-Mode can be combined with Color Mode.
Preparations
1. Connect one of the probes, to the Ultrasound system’s
left-most probe connector.
2. Turn ON the Ultrasound system.
The 2D Mode window is displayed (default mode).
3. Press MM on the Operating Panel to bring up an M-Mode
picture on the screen.
4. Use the trackball to position the cursor over the required
area of the image.
Using M-Mode
Conventional M-Mode
1. To access M-Mode from any other scan mode, press MM on
the control panel.
2. Use the trackball to position the cursor over the required
area of the image.
3. Press Freeze.
4. Use the trackball to scroll through the data acquired.
Anatomical M-Mode
1. In M-Mode or 2D-Mode Freeze, press AMM on the Touch
panel.
NOTE: Anatomical M-Mode can also be used with previously
acquired digitally stored 2D images. More than one heart
cycle should be stored if performing M-Mode in post
processing.
2. Use the trackball (assigned function: Pos) to position the
cursor over the required area of the image.