Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
General Procedures and Functional Checks
4-16 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
3. Press Trackball to allow free rotation of the solid full-length
cursor line throughout the 2D image (trackball assigned
function: Angle).
4. Rotate the solid cursor line to the desired direction.
Curved Anatomical M-Mode
1. In M-Mode, press Curved AMM.
2. Use the trackball (assigned function: Pos) to position the
starting point of the time motion curve.
3. Press Select to anchor the starting point of the time motion
curve.
4. Use the trackball to position the next point of the time motion
curve.
5. Press Select to anchor the point of the time motion curve.
6. Repeat step 4 and 5 up to draw a complete time motion
curve.
NOTE: The time motion curve can be edited by following the curve
back to the desired point and redraw.
7. On the last point, press Select twice to terminate the curve.
NOTE: To edit the time motion curve, select a point, move it to a
new position and press Select.
Optimizing M-Mode
The use of preset gives optimum performance with minimum
adjustment. If necessary, the following controls can be adjusted
to further optimize the M-Mode display:
NOTE: Refer to ‘Creating presets’ on page 4-8 about creating presets.
• Adjust Horizontal sweep to optimize the display resolution.
• Adjust Gain and TGC controls to adjust the range to be
imaged.
• Use the Frequency (move to higher frequencies) or the
Frame rate control (move to lower frame rate) to increase
resolution in image.
• Adjust Dynamic range to optimize the useful range of
incoming echoes to the available grayscale.
• Adjust Compress to further optimize the display.
• Adjust Reject to reduce noise while taking care not to
eliminate significant low-level diagnostic information.