Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Diagnostics/Troubleshooting
7-8 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Troubleshooting
XY Lock is not working
Follow the steps below to resolve this issue:
1. Run the XYZ Test Program and select Fix Lock.
2. If the Lock still does not respond when pushing the buttons
in front of the UI, release the Lock manually.
3. Remove the Park Lock and adjust it manually.
Related information:
• ‘Test the motor controller’ on page 7-10
XY Brake Motors Troubleshooting
NOTE: Cables are not labeled at the XY controller, so right side and left
side motors are not defined.
When troubleshooting the XY brake motors, keep this in mind:
• The rear motors are #1 and #3.
• The front motors are #2 and #4.
• Motor #1 and #2 are on the same side.
• Motor #3 and #4 are on the same side (opposite to motor #1
and #2).
If you are in a silent environment, it may be possible to hear
which motor is activated during the test.
Z Movement fails
• If it is impossible to lower the Top Console but moving it
upwards works OK, one of the control switches is stuck in
the “Up” position.
• If you have to help the Top Console up when moving
upwards, but the motor assistance work OK when lowering
the Top Console, the gas spring inside the Z Mechanism is
failing: