Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Warnings and important information
Vivid E80/E90/E95 – Service Manual 8-3
GC091052 Rev. 3
NOTE: Use an ESD compatible work space or the ESD-kit during parts
replacement.
Returning/shipping probes and repair parts
Equipment being returned must be clean and free of blood and
other infectious substances.
GE policy states that body fluids must be properly removed from
any part or equipment prior to shipment. GE employees, as well
as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no
circumstance should a part or equipment with visible body fluids
be taken or shipped from a clinic or site (for example, body coils
or an ultrasound probe). The purpose of the regulation is to
protect employees in the transportation industry, as well as the
people who will receive or open this package.
WARNING
The waste of electrical and electronic equipment must not be
disposed as unsorted municipal waste and must be collected
separately.
Please contact the manufacturer or other authorized disposal
company to decommission your equipment.
WARNING
Spare Part order for Vivid E80/E90/E95 VET:
The same spare parts used in the current Vivid E80/E90/E95
consoles can be used in the veterinary consoles.
When ordering any of the following, the Vet label(s) MUST be
ordered as they need to be applied to the new part:
• Side Covers - require label P/N: 5447716
• Back Cover - require label P/N: GB200182
• Probes - require label P/N: 5454608
CAUTION
When performing replacement procedures within a veterinary
environment, return any used/unused spare parts with the
purple scrap marking label (regardless of the actual condition
of the part) and add a description on the label stating that the
items were removed from a veterinary environment.