Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Replacement Procedures
8-40 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Preparations
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Rotary knobs below the Touch Panel and the
Volume button.
5. Remove Operating Panel, Upper.
Remove the Operating Panel, Lower
Follow these steps to remove the Operating Panel:
1. Pull the Alphanumeric keyboard to its extended position to
get access to the screws in the next step.
2. The screws are available from the underside of the
Operating Panel tray. Unscrew and remove the four screws
used for fixing the Operating Panel assembly to the
Operating Panel tray.
3. Loosen the screws for the cables grounding on the
Bulkhead Bracket and move the OP Cables away from the
bracket.
4. On the Bulkhead Bracket, loosen the ESD wire from the
Lower Panel.
5. Pull and lift the Operating Panel assembly up and away. Be
careful to not destroy the fingers on each side of the
Operating Panel, Lower.
6. Store it on an ESD safe place.
Install the Operating Panel, Lower
Follow these steps to install the Operating Panel, Lower:
1. Carefully, slide in the Operating Panel. Lower.
Be careful with the fingers which have to be placed under
the side walls of the UI Frame, Upper.
2. On the Bulkhead Bracket, fasten the ESD wire from the
Lower Panel.
3. Attach the cables to the cable grounding brackets/points, as
illustrated below. Since this feature is for the EMI
compatibility the braid have to be exposed under the bracket
and the cables have to be well locked by the bracket.
4. Install the four fixing screws that fix the Operating Panel
assembly to the Operating Panel tray (2 pc. M4x25 nearest