8-56 Vivid E80/E90/E95Service Manual
GC091052
Rev. 3
The lower left-most screw is M6 x 30, Torque: 8.5 Nm.
The three other lower screws are M6 x 25, Torque: 8.5
Nm. Two of the screws are also fixing the Cable Clamp
3. Install the Operating Panel, Lower.
4. Install the Bulkhead Board.
5. Install the Operating Panel, Upper.
6. Install the Operating Panel Knobs.
Handle, Left Top / Handle Right Top, replacement
Read and follow
Preparations
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
WARNING
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
WARNING
DO NOT touch any boards with integrated circuits prior to
taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised
ESD connection point located on the rear of the Ultrasound
system (near the power connector).
Follow general guidelines for handling of electrostatic sensitive
equipment.