Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Replacement Procedures
8-64 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
3. Disconnect all probes and external I/O cabling.
4. Remove the LCD Rear Cover.
5. Remove the LCD Monitor assembly.
6. Remove the LCD Arm assembly.
7. Remove the Rotary buttons below the Touch screen and the
Volume button.
8. Remove the Operating Panel, Upper.
9. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
10. Remove the Bulkhead Board.
11. Remove the Bulkhead Plate.
12. Remove the Upper Frame Assembly.
13. Remove the Lower Frame Assembly.
Remove the XY Mechanism
Follow this procedure to remove the XY Mechanism:
1. From the rear side of the Ultrasound system, remove the
four (4x) fixing screws.
2. Remove the XY mechanism.
Install the XY Mechanism
Follow this procedure to install the XY Mechanism:
1. Position the XY mechanism so it aligns with the holes for the
four fixing screws.
2. Install the four fixing screws.
3. Install the Lower Frame Assembly.
NOTE: Ensure that Bumper Z enters the slide at the back, before
tighten the screws.
4. Install the Upper Frame Assembly.
5. Install the Bulkhead Plate.
6. Install the Bulkhead Board.
7. Install the Operating Panel, Lower.
8. Install the Operating Panel, Upper.
9. Install the LCD Arm assembly.
10. Install the LCD Monitor assembly.
11. Install the LCD Rear Cover.