Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Replacement Procedures
8-82 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Rod Assembly removal
1. Remove the screw that ties the rod assembly to the pedal
mechanism. This requires unscrewing the nuts in the back
while holding the screw on the front.
2. Remove the nut on the left and the casters on the right that
tie to the rod.
Rod Assembly installation
1. Position the pedal mechanism and install the top four fixing
screws (Torque: 10 Nm (7.4 lb.-ft.)).
2. Install the two bottom screws.
3. Install the screw that ties the rod assembly to the pedal
mechanism. You need to hand tighten the nut that attaches
the rod to the wheel.
4. Connect the rods by snapping on the ball joints.
Table 8-3: Rod Assembly installation
Step Corresponding Graphic
1. All the washers must be placed in the
sequence as shown.
• Place the screw (1) through the
small, flat washer, followed with a
large, flat washer.
• The spacer (2) will seat between the
two large, flat washers.
• Place the screw, washers and
spacer through the Brake Rod
Assembly and the Pedal Assembly
Mechanism. Follow with the other
large, flat washer.
• Tighten the first 10 mm nut (3) to the
screw, washers and spacer.
• Install the 10 mm locking nut (4) and
tighten.
NOTE: An 8 mm wrench is
required when tightening the
pivots to the actuator lever on
the front castors.
• Check ALL other nuts (5) on the
Brake Rod Assembly to ensure ALL
are tight.