Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Replacement Procedures
8-98 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Warnings (continued)
Preparations
When preparing for the replacement, you must perform the
following steps:
1. Record the Ultrasound system’s TCPIP settings and
installed Option strings.
2. Export the Patient Archive images to a server or to external
media.
3. Back up the Report Archive, User-Defined Configuration
(Customer Presets), and Service settings.
You will perform a Restore after you have installed the new
BEP and loaded the correct software.
4. Collect the error logs and record information on why the
BEP must be changed.
5. Reload the system software from removable media. Select
No when asked: Keep User Data and Patient Archive?
6. Power down the Ultrasound system.
7. Disconnect the Mains Power Cable from the wall outlet.
8. Disconnect all probes.
9. Remove the Side Covers.
10. Remove the Top Cover.
11. Remove the B/W Printer.
12. For easier access in the next steps, remove the Front Cover.
WARNING
Before you dispose of the hard drive, or return the BEP to the
local parts organization, make sure you remove ALL PATIENT
DATA from the hard drive, given that the hard drive is still
functional. In some countries, you may be required to delete all
software from the disk before returning the hard drive to the
parts warehouse. Follow your local policies.