Product Manual
Table Of Contents
- Vivid™ E80 / Vivid™ E90 / Vivid™ E95
- Table of Contents
- Chapter 1
- Introduction
- Chapter 2
- Site Preparations
- General Ultrasound system requirements
- Facility needs
- Environmental Dangers
- Chapter 3
- System Setup
- Setup reminders
- Receiving and unpacking the equipment
- Packing materials - recycling information
- Preparing for setup
- Completing the setup
- Configuration
- Connectivity overview
- Connectivity setup
- Product Locator Installation Card
- Chapter 4
- General Procedures and Functional Checks
- Chapter 5
- Components and Functions (Theory)
- Software overview
- InSite ExC
- Vivid E80/E90/E95 overview
- Top Console with LCD monitor and Operating Panel
- Main Console
- Air Flow control
- Casters and brakes
- Front End Processor (FEP)
- Back End Processor (BEP)
- Power distribution
- Input and Output (I/O) modules
- Restart Vivid E80/E90/E95 after diagnostics
- Chapter 6
- Service Adjustments
- Chapter 7
- Diagnostics/Troubleshooting
- Chapter 8
- Replacement Procedures
- Warnings and important information
- Loading the software
- Replacing covers and bumpers
- LCD Monitor and LCD Arm parts replacement
- Upper OP Panel/Touch Panel Assembly replacement
- Lower Operating Panel Parts replacement
- A/N Keyboard parts replacement
- Other Top Console Parts replacement
- Replacing XYZ Parts
- Main Console parts replacement
- Casters and Brakes replacement
- Front End Processor (FEP) / Card Cage parts replacement
- Back End Processor (BEP) parts replacement
- Main Power Supply replacement
- Patient I/O assembly replacement
- Peripherals replacement
- Cables replacement
- Chapter 9
- Renewal Parts
- List of Abbreviations
- Software for Vivid E80/E90/E95
- Covers and Bumpers
- Top Console parts
- XYZ parts
- Main Console parts
- Casters (wheels) parts
- Front End Processor (FEP) Card Rack parts
- Back End Processor (BEP) parts
- Main Power Supply
- I/O modules parts
- Peripherals for use with Vivid E80/ E90/E95
- Mains power cables - Vivid E80/E90/ E95
- Internal Cables - Vivid E80/E90/E95
- ECG cables - Vivid E80/E90/E95
- Physio TX parts
- Probes for Vivid E80/E90/E95
- Options - Vivid E80/E90/E95
- Chapter 10
- Care and Maintenance
- Index
Replacement Procedures
8-100 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Back End Processor (BEP) installation procedure
Follow these steps to install the BEP:
1. Slide the BEP into the left side of the chassis frame.
2. Install the two (2x) hex key screws at the inside base of the
BEP.
3. Reach inside the BEP and connect the PCIe cable.
NOTE: If you don’t reconnect the PCIe cable to the backplane, the
Ultrasound system will power up, but will not scan. The BEP
will launch into simulator mode.
4. Install the BEP Cover.
5. Replace the Printer Bracket at the top of the BEP.
Be sure the lip, on the underside of the bracket, hooks on
the edge of the Card Rack, and the 3 tabs insert into the
slots on the top of the BEP frame. The lip “clamps” the Card
Rack and BEP together. This is a tight fit.
6. Lower the Lower Column Cover.
7. Position the Front Cover to engage the Column Cover Stop
Tabs.
8. Latch the two latches that clamp the Printer Bracket to the
top of the BEP.
9. Slide the B&W printer into the Printer Bracket and connect
the cables to the back of the printer.
10. Tighten the Printer Bracket Wing Nut to secure the printer.
11. Connect all I/O cables at top and/or face of the BEP,
including the printer cables.
12. Install the Top Cover (if removed), the Front cover (if
removed) and the side covers.
13. If applicable, install the correct software on the new BEP.
14. Enter the Ultrasound system’s TCPIP settings and Option
strings.
CAUTION
Be careful not to pinch any of the cables.
CAUTION