General Ultrasound system requirements
Vivid E80/E90/E95Service Manual 2-5
GC091052 Rev. 3
Electrical requirements for the Vivid E80/E90/E95
In the table below, the electrical specifications for the Vivid E80/
E90/E95 includes monitor and on-board peripherals.
The specifications apply for the following models:
GC000220 - Vivid E95
GC000230 - Vivid E90
GC000240 - Vivid E80
The current drain will vary depending on the mains voltage.
At 230 VAC the current may be up to 3 A.
At 100 VAC the current may be up to 7 A.
Site circuit breaker
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
Site power outlets
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
Table 2-3: Electrical specifications
VOLTAGE TOLERANCES POWER
CONSUMPTION
FREQUENCY
100-240 VAC ±10% 700 W / 770 VA 50/60 Hz
CAUTION
Power outage may occur. The Vivid E80/E90/E95 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you do not
have any other equipment operating on the same circuit.