4-58 Section 4-7 - Peripheral Checks
Figure 4-54 Extender
18.)Hold the cast base firmly and place it on the Extender as shown in the images below
Figure 4-55 Pull the cables and place the Extender
Assemble Extender as shown and pull
Extender
Fasten 4 screws, 4 spring
washers and 4 flat washers
all the cables through the Extender
Pull all the cables through (Speaker cable,
LCD cable, OPIO cable) Keyboard base
Place the Keyboard base on the Extender