Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-49
5-12-2 Error Logs
When the Error Logs page is selected, different log viewing options are available.
Log Viewer is displayed in a separate window.
5-12-3 Diagnostics
The Diagnostic page uses a web-controlled user interface to provide access to common service
components and perform diagnostics.
Non-Interactive: The tests are performed without the user’s intervention.
Interactive: The user is required to perform an operation on the ultrasound unit in order for the test
to be completed successfully. This option is not applicable when used remotely.
Figure 5-26 Common Service Desktop - Error Logs
Figure 5-27 Common Service Desktop - Diagnostics