5-54 Section 5-13 - Service Page
5-13-3-8 Keyboard Layout
To change the keyboard layout to different languages.
Operation see:
Section 6-4 "Modification of Keyboard Layout" on page 6-3.
NOTE: Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
see:
Section 8-38 "Replacement of AN KEY TOP" on page 8-83
5-13-3-9 Monitor Test
Select the MONITOR TEST
button to perform color calibration.
Operation see:
Section 6-3 "LCD Monitor Adjustment" on page 6-2.
5-13-3-10 Load Application Settings
If the Tune version of the Application presets does not match the Application Software version, it is
probably that there are adverse affects on image quality (e.g., after reloading an old “Full Backup”).
Select the LOAD APPLICATION SETTINGS
button on the “Service Tools” page to load proper
Application Settings (image presets), adapted for the used system software version.
1.) Select the LOAD APPLICATION SETTINGS
button on the “Service Tools” page (Figure 5-33 on
page 5-52
).
2.) Choose the media and then click LOAD
.
3.) Select the desired file and then click OK
.
Load procedure starts immediately including a reboot of the system.
4.) If warning message is displayed, confirm it with OK
and then load appropriate Application Settings
(perform loading procedure as described in steps above).
NOTICE
!! NOTICE:
When reloading these Application Settings, any existing User Programs, 3D/4D Programs and Auto
Text remain unaffected!
Figure 5-35 Application Settings Load
NOTICE
!! NOTICE:
If the ID of the Application Setting is not valid for the currently installed Application Software version,
a warning message “Application Settings from selected backup are not compatible with current system.
Application Settings have not be loaded.” appears during boot up sequence.
Choose media
and
then click [Load]