Chapter 5 - Components and Functions (Theory) 5-63
4.) In the displayed menus select [Exit] and then [Exit/Reboot].
5.) Remove the System DVD from the DVD/CD+R/RW Drive drive.
6.) Please wait until all processes are finished and the final “DOS” window (Figure 5-43 below) appear.
7.) Press A key to activate the Ultrasound Application Software [UIS].
Figure 5-42 Exit -> Exit/Reboot
NOTICE
The system boots with windows. An automated process was developed to check disk, assign an unique
computer name and matching drive letters. The system will reboot at least 2 times!
Figure 5-43 press “A” key to start UISApp