Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
5-64 Section 5-15 - LINUX - Voluson Maintenance System
8.) If following message appears, skip the “BT warning” with CONTINUE.
NOTE: Options will be recovered at a later step using “Full Backup” process.
9.) Customize the system as described in
Section 5-15-7 on page 5-75.
5-15-5-3 Initialize
This function performs operations which result in complete data loss of all files on C:\ and D:\
partition. The rescue image stored on the Rescue (R:) partition of the hard disk is replaced by the
image from the DVD.
5-15-5-4 NewDisk
NOTE: This function is ONLY available when booting from System DVD.
This function performs operations which result in complete data loss of all files on the entire hard
disk. It re-initializes the hard disk using partition images on the System DVD (setup to factory default).
Figure 5-44 Skip the BT warning with CONTINUE
WARNINGWARNING
All data on D:\ (User) partition (e.g., Patient Archive images, Service Database,
Measurement results, Patient Data and Reports) will be deleted and re-initialized.
NOTICE
If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 13.0.0 or higher, a new software specific”Permanent key” is required.
NOTICE
The [Initialize] function should only be performed if NO System DVD is available!
If you have a System DVD, it is highly recommended to use the [New Disk] function
(see:
Section 5-15-5-4 on page 5-64), which is only available when booting from the System DVD.
WARNINGWARNING
COMPLETE DATA LOSS ON THE ENTIRE HARD DISK. Also data on D:\ [User] partition
(e.g., Patient Archive images, Service Database, Measurement results, Patient data and
Reports) will be lost and can not be carried over to the new hard disk.