Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 6 - Service Adjustments 6-3
Section 6-4
Modification of Keyboard Layout
NOTE: Configuring the layout of the keyboard is only useful by changing the concerned keys also.
see: Section 8-38 "Replacement of AN KEY TOP" on page 8-83
6-4-1 Setup the Voluson® P8 / Voluson® P6 Keyboard Language Layout
1.) Press the F2
key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the SERVICE. The
“password window” appears automaticaly.
4.) Enter the password
SHE and click the ACCEPT
button to display the Service Tools window.
5.) Click the KEYBOARD LAYOUT button.
Figure 6-2 Service Tool window
[Keyboard Layout]
button