Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-21
7-7-1 System does not boot up
Figure 7-22 System does not Power On / Boot Up
System doesn’t boot
Is Standby Switch
illuminated?
Press Standby Switch till it
changes to green System
starts booting
Is Standby Switch
orange?
Is there any picture on
the monitor ?
Other error. Please contack
the service engineer
Is there any picture on
the monitor ?
- Power Cord is connected to Mains
- Power Switch is turned on
(on the backside of the system)
Verify that
Press Standby Switch for
approximately 5 sec.
untill system is down
Check cables between
Console and Monitor
No
No
No
No
Yes
Yes
Yes
Yes