Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
7-32 Section 7-8 - Error Messages
cannot read file restart the system
Cannot import volume data to 3D dll
Action: Load volume files from other storage medium. Call technical support
Load volume files from other storage medium.
Can't add chosen files to the virtual folders! restart the system
Can't create hardware configuration index file in D:\\SERVICE
Action: restart the system
restart the system
Can't detect PCI 9054 restart the system
Can't find %s() dll function in Mizar_IntensityMeasureData.dll restart the system
Can't import disc session.
Action: use a new CDR to write data to CD, if not ok call technical support
use a new DVD/CD+R to write data to DVD/CD
Can't load Mizar_IntensityMeasureData.dll restart the system
Can't open: .......... restart the system
Can't open: patterndat.bin Action: restart the system restart the system
Can't open: resampcoef.bin Action: restart the system restart the system
CHA PulseInversion flag for this probe is not set
Action: restart the system
restart the system
Cine2D_CtrlBlock::vSetState: value of m_GIP_eCineType undefined!! restart the system
Cine2D_CtrlBlock::vSetState: value of m_pGIP_eR_W_Mode undefined!! restart the system
CKV_Wrapper: initialization failed restart the system
Communication thread is dead! Action: Restart the system.
Check connection from System to Recorder, Recorder has power and is on.
restart the system; check connection from system to VCR,
VCR has power and is on
ConManager - RC_ConstructGOPs: Memory allocation failed restart the system
Corrupt CVIE parameter file restart the system
CPI PulseInversion flag for this probe is not set
Action: restart the system
restart the system
CPI-mode is only with curved probe Action: Contact technical support restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for ColorDMA_Active restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for EndOfFrameInt restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for LongLineMemFIFO restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for Mode_4D restart the system
CRI: mismatch in steering angle header info
Action: restart the system
restart the system
CtrlMvGrpRepresentations::isInWriteMode not supported for mode nr xx restart the system
CVIE parameter file could not be created restart the system
CW Archive::not implemented. Action: restart the system
restart the system
Error Messages Actions