Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-33
CW quant TxFreq out of range
restart the system
CW_BaseLinePos Out Of Range
Press ok and save this user-setting once again.
CW_Gain Out Of Range Press ok and save this user-setting once again.
CW-hardware doesn't support
pencil probe + CWHardware not available, restart the system
CW-TxFrequency == 0
restart the system
D:\\SERVICE\\ConfIndexFile is corrupted Action: restart the system restart the system
Data are not saved or sent to DICOM Server! restart the system
dDistanceFocusReferenceBF_ToProbeSurface out of valid range Action: restart the
system
restart the system
Delete error (Backup data could not be deleted.)
check storage destination of Full Backup
(e.g., DVD/CD, insufficient rights on target Network drive,
write protection on USB-drive)
Density Out Of Range Press ok and save this user-setting once again.
Device is not a DVD/CD rec. reconnect the device
Different software error (Backup data was made on another system with a different
software version.)
This backup cannot be restored by the user.
Disc is full! use a new CDR for writing data to CD
Display:Rect Region fails restart the system
Do you really want to delete?
Do you really want to format this DVD?
done is low! restart the system
Doppler-Txfrequency out of range restart the system
Downgrade error (Backup data was made with a software version higher than the
installed version.)
load appropriate backup for installed version
DVD Recorder communication error! Check Recorder cables and try again.
Make sure DVD Recorder is connected to US machine.
check DVD connections, check if DVD is switched on
DVD Recorder not properly connected or turned off!
Check connection from US machine to Recorder, Recorder has power and is on.
check DVD connections, check if DVD is switched on
DVD Recorder timeout error! Check Recorder cables, DVD,... and try again. check DVD connections, check if DVD is switched on
ECG Module not connected. connect ECG Module
ECG_Consumer: memory allocation failed
restart the system, reinstall UISAPP, contact online
center(OLC)
Electronic user manual not installed. Please install. install Electronic User Manual (EUM) and try it again
End Bandwidth too big restart the system
End ET too big restart the system
End frequency too big restart the system