Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
8-20 Section 8-8 - Replacement or Activation of Options
8-8-1 Operation for activating Options
1.) Press the F2
key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the OPTIONS tab (see:
Figure 8-9 on page 8-11.)
8-8-1-1 Operation for installing a “Demo Key” or a “Permanent Key”:
1.) Position the cursor inside the input field desired and press the
right/left trackball
key.
2.) If one exists, clear/edit the current key code.
3.) Enter the encrypted serial code with the keyboard and click on SUBMIT. (Code will be checked.)
4.) Click the SAVE&EXIT button.
NOTE: After activating a key code, restart (turn off and on) the Voluson® P8 / Voluson® P6 system.