Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-27
Section 8-12
Replacement of the SPEAKER
8-12-1 Manpower
One person, 30 minutes
8-12-2 Tools
Philips screwdriver
8-12-3 Preparations
1.) Power Off/Shutdown the system as described in
4-3-2 on page 4-4.
8-12-4 Removal Procedure
1.) Remove the MAIN OPIO and SUB OPIO. Refer to the
Section 8-11 on page 8-24 OPIO.
2.) Unscrew each 2 screws of SPEAKER ASSY L, R. Refer to the figure below.
Figure 8-29 Unscrewing 2 screws to separate SPEAKER ASSY
Figure 8-28 Unscrewing 6 screws to separate SPEAKER ASSY