Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
8-46 Section 8-19 - Replacement of the Trackball
4.) Disconnect the TRACKBALL CABLE from the OPIO HUB PWA. Refer to the figure below.
5.) Unscrew 2 screws and then pull out the TRACKBALL from the MAIN OPIO. Refer to the figure
below.
Figure 8-55 Disconnecting the TRACKBALL CABLE
Figure 8-56 Unscrewing 2 screws and pulling out the TRACKBALL