DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-81
Section 8-37
Replacement of the OPIO USB Sub PWA
8-37-1 Manpower
One person, 30 minutes
8-37-2 Tools
Philips screwdriver
8-37-3 Preparations
1.) Power Off/Shutdown the system as described in
4-3-2 on page 4-4.
2.) Remove the rear cover as described in Section 8-11 on page 8-24.
8-37-4 Removal Procedure
1.) Pull 2 cables from Main Keyboard ASSY, and prepare the Sub Keyboard ASSY.
Figure 8-97 Pulling 2 cables
Figure 8-98 Unscrewing 5 screws