DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-105
Section 8-43
Functional Checks after FRU replacement (Debrief Guidelines)
1.) Perform required Functional and/or Leakage Current tests based upon the FRU being replaced.
2.) Clarify the scripts to debrief a Service Dispatch using your pole's dispatch tool.
Use this script in the Service Comments when debriefing a Service Dispatch.
For a replaced FRU, use the following debrief script:
Proprietary Service Manual, Voluson® P8 / Voluson® P6, Revision Voluson® P8 / Voluson® P6,
Section [fill in appropriate section]. Equipment passed all required tests and is ready for use.
And, if testing leaking current, use the following text.
Proprietary Service Manual, Voluson® P8 / Voluson® P6, Revision Voluson® P8 / Voluson® P6,
Section [fill in appropriate section]. Leakage Current meets allowable limits. Equipment passed
all required tests and is ready for us.