Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-109
6.) Pull out BF64/PIR Boards.
Figure 8-139 BF64/PIR Boards
7.) Unscrew 6 screws and replace BF64(p/n:5396937) to BF64C Boards which have CWD Circuits.
Figure 8-140 BF64C Board
8-44-5 Installation Procedure
1.) Install the new board assy in the reverse order of removal procedure.
2.) Perform:
- Functional Tests
.
Unscrew 6 screws