DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
8-110 Section 8-45 - Replacement of ECG Kit assy
Section 8-45
Replacement of ECG Kit assy
8-45-1 Manpower
One person, 10 minutes.
8-45-2 Tools
Philips screwdriver
8-45-3 Preparations
Power Off/Shutdown the system as described in
4-3-2 on page 4-4.
8-45-4 Removal Procedure
1.) Remove 4 probe holders.
2.) Unscrew 9 numbers of fasteners below the keyboard base.
Figure 8-141 Keyboard base