Chapter 10 - Care & Maintenance 10-13
10-6-7-4 Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when
transporting the system with the probe attached, or during the disinfection process.
Inadequate handling may lead to dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
NOTE: Operation instructions are supplied with each RIC-Holder (KTZ225469).
Endocavity Soaking Cup Kit
is available at GE Accessory
(Order Nr. : E8385TA )
To guarantee that the sensitive probe head does not hit the bottom of
the container,place the probe in the RIC Holder during the disinfection
procedure.
OKOK
Disinfection fluid
RIC-Holder
Endocavity soaking cup kit
RIC-Holder
Container
RIC-Holder