Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
10-18 Section 10-8 - Electrical Safety Tests
NOTE: * Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV
** Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table
19.5DV.1 *** Measurement limits per IEC 60601-1 Ed 2. Medical Electrical Equipment Safety
Standards, Table IV
**** Isolation between Mains and Applied Limits refers to the measurement of leakage current
flow which would flow from mains to patient if the patient were to come into contact with mains
voltage.
*****Measurement limits per IEC 60601-1 Ed 3. Medical Electrical Equipment Safety Standards, Table 3
******Measurement limits per IEC 60601-1 Ed 3. Medical Electrical Equipment Safety Standards, Table
4
The following tests are performed at the factory and should be performed at the customer site as well.
- Grounding Continuity
- Earth Leakage Current
- Enclosure Leakage Current
- Patient Leakage Current
All measurements should be made with an electrical safety analyzer.