10-22 Section 10-8 - Electrical Safety Tests
ii)Others
Hot : Brown
Neutral : Blue
Ground : Green-yellow
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool
compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
When using a test instrument, its power plug may be inserted into the wall outlet and the equipment
under test is plugged into the receptacle on the panel of the meter. This places the meter in the
grounding conductor and the current flowing from the case to ground will be indicated in any of the
current ranges. The maximum allowable limit for chassis source leakage is shown in
Table 10-14 on
page 10-15.