Product Manual

GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-35
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Type CF Patient Auxiliary Current Test - ECG Lead to Probes
Type BF Patient Leakage Current Test - Mains to Probe
Probe (This test should be done for all probes)
Type CF Patient Leakage Current Test - Mains to ECG leads
ECG leads connected all together
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.1mA
Closed Normal Open 0.5mA
Closed Reversed Closed 0.1mA
Closed Reversed Open 0.5mA
Open Normal Closed 0.5mA
Open Reversed Closed 0.5mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.01mA
Closed Normal Open 0.05mA
Closed Reversed Closed 0.01mA
Closed Reversed Open 0.05mA
Open Normal Closed 0.05mA
Open Reversed Closed 0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 5mA
Closed Reversed Closed 5mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.05mA
Closed Reversed Closed 0.05mA