10-36 Section 10-9 - When There's Too Much Leakage Current...
Total Patient Leakage Current - Type CF probe to Ground
Total Patient Leakage Current - Mains to Type CF probes
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.05mA
Closed Normal Open 0.1mA
Closed Reversed Closed 0.05mA
Closed Reversed Open 0.1mA
Open Normal Closed 0.1mA
Open Reversed Closed 0.1mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.1mA
Closed Reversed Closed 0.1mA