DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-3
3-2-2-4 Brake Pedal Operation
3-2-3 Safety Reminders
WARNINGWARNING
!! WARNING:
REMEMBER: If the front wheel brakes are engaged for transportation, release brake
pedals (brakes on front wheels under the foot rest) to disengage the lock.
DANGER
!! DANGER:
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT
TOUCH THE UNIT!
CAUTION
!! CAUTION:
Two people should unpack the unit because of its weight.
Two people are required whenever a part weighing 16kg (35 lb.) or more must be lifted.
CAUTION
!! CAUTION:
If the unit is very cold or hot, do NOT turn on its power until it has had sufficient time to
acclimate to its operating environment.
CAUTION
!! CAUTION:
To prevent electrical shock, connect the unit to a properly grounded power outlet.
DO NOT use a three to two prong adapter. This defeats safety grounding.
CAUTION
!! CAUTION:
DO NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
CAUTION
!! CAUTION:
DO NOT use a 20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord.
This unit requires a dedicated 16 A circuit.
CAUTION
!! CAUTION:
DO NOT operate this unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling.
(When covers are removed, EMI may be present).
CAUTION
!! CAUTION:
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the Voluson® P8 /
Voluson® P6. Keep manuals near the unit for reference.
CAUTION
!! CAUTION:
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Voluson® P8 / Voluson® P6 is within
FDA limitations, avoid unnecessary exposure.
Ultrasound energy can produce heat and mechanical damage.
Figure 3-2 Environmental Labels