Product Manual

GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) BASIC SERVICE MANUAL
3-8 Section 3-4 - Preparing for Set Up
Section 3-4
Preparing for Set Up
3-4-1 Verify Customer Order
1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have
been received. Compare all items listed on the packing slip (delivery note) with those received.
2.) Visually inspect the system components using the following checklist.
NOTE: Report any items that are missing, back-ordered, or damaged, to your Voluson® P8 / Voluson® P6
sales representative. The contact address is shown in
Contact Information on page 1-16.
3-4-2 EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
NOTICE
!! NOTICE:
It is recommended to keep and store the shipping carton and all other packing materials (including the
support foams, anti-static plastic cover, etc.), in case the unit has to be moved to a different location.
Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase.
Table 3-6 Damage Inspection Checklist - Voluson® P8 / Voluson® P6 System
Step Item Recommended Procedure
1
Main label Enter Serial Number: __________________________ (printed on main label, see:
Figure 1-2 on page 1-11)
2
Console Verify that the system is switched OFF and unplugged. Clean the console and control panel.
3
Control Console
Physically inspect the control console for missing or damaged items.
After switching on the system, verify the proper illumination of all the control panel buttons.
4
Probes
Check all probes for wear and tear on the lens, cable, and connector. Look for bent or damaged pins on the
connector and in the connector socket on the unit. Verify that the EMI fingers around the probe connector
socket housing are intact. Check the probe locking mechanism and probe switch.
5
LCD Display
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded cloth. Inspect the monitor
for scratches and raster burn.
6
Fans Verify that the system’s cooling fans and peripheral fans are operating.
7
Rear Panel
Check the rear panel connectors for bent pins, loose connections and loose or missing hardware. Screw all
the cable connectors tightly to the connector sockets on the panel. Verify that the labeling is in good condition.
8
Covers
Check that all screws are tightly secured in place, that there are no dents or scratches and that no internal
parts are exposed.
9
Peripherals
Check and clean the peripherals in accordance with the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
10
Power Cord
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity. Replace the power cord, as required.
WARNINGWARNING
!! WARNING:
CONNECTING A Voluson® P8 / Voluson® P6 SCANNER TO THE WRONG VOLTAGE
LEVEL WILL MOST LIKELY DESTROY THE SCANNER.
b