Chapter 3 - Setup Instructions 3-33
away.
13.)Connect the Bluetooth Adapter (Belkin or DBT-122) to USB port E (see: Figure 3-8 on page 3-16)
on the rear side.
14.)After plugging in the Bluetooth Adapter, wait (~ 3 minutes) until the "Found New Hardware Wizard"
has installed the Bluetooth Dongle.
15.)When the "InstallShield Wizard Completed dialog appears, click on FINISH
.
NOTE: If dialog "You must restart your system..." appears, click on YES
to restart the system.
16.)Remove the DVD, close all open windows and restart the system (turn off and on the system).
3-7-3-2 Configuration of the Bluetooth Software
1.) After the system has restarted, connect the USB "standard GE" service dongle on the front USB
connector.
2.) Enter the "Standard GE Revolving" password, click ENABLE
.
3.) Press the ALT + F4 (= "Eject") key again to close the Ultrasound Application.
Figure 3-23 “Bluetooth device not found” - DO NOT click any button
NOTICE
!! NOTICE:
ONLY when using “obsolete” DBT-120 Bluetooth adapter:
If prompted, enter the 32-digit security code (8 digits per field), which is located on the back flap of the
driver CD-cover and then click OK
.
Figure 3-24 Finish installation and restart the system
If the “Bluetooth device not found”
dialog box obstructs the view,