- General Electric Water Pump User Manual

GE Analytical Instruments, Inc. ©2005H Page 12 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
CONSTANT VALUE
NAME
CONSTANT VALUE AFTER
UPGRADE
CONSTANT VALUE PRIOR
TO UPGRADE
ANALOG MAX:
Same as prior value
TOTAL ACID VOL: 150
TOTAL OXID VOL: 150
ALARM #1 TYPE:
Same as prior value
ALARM #2 TYPE:
Same as prior value
ALARM #1 LEVEL:
Same as prior value
ALARM #2 LEVEL:
Same as prior value
LAMP MAX DAYS: 182
HOSE MAX DAYS: 365
OXID MAX DAYS: 90
HYSTERESIS CNT: 40
TC/IC OFFSET (60Hz):
Same as prior value
SLOPE PPM/HOUR: 5000
S1 TOC ACID FLOW: 0.2
S1 TOC OXID FLOW: 0
S2 TOC ACID FLOW: 0.4
S2 TOC OXID FLOW: 0.2
S3 TOC ACID FLOW: 0.75
S3 TOC OXID FLOW: 0.5
S4 TOC ACID FLOW: 0.75
S4 TOC OXID FLOW:
2.5 or 2.0
same as prior value
REFILL DELAY: 15
SERIAL NUMBER:
Same as prior value
TC TEMP CORR A: 2.03149
TC TEMP CORR B: -0.0593627
TC TEMP CORR C: 0.00101571
TC TEMP CORR D: -5.7161E-06
IC TEMP CORR A: 2.57516
IC TEMP CORR B: -0.0975902
IC TEMP CORR C: 0.00194737
IC TEMP CORR D: -1.34084E-05
DAC RANGE MIN:
Same as prior value
DAC RANGE MAX:
Same as prior value
PRINTOUT INTERVAL: 0
SAMPLE STEPZERO: 100
SAMPLE FLOWRATE:
55 or 71
same as prior value
SYRINGE MAX POINT: 31000
SYRINGE MAX RANGE: 3100
SYRINGE ZERO RANGE: 150
REAGENT FLUSH REPS: 2
NUMBER OF STREAMS: 4
AC FREQ. SETTING:
Same as prior value