Operating Manual
www.ge.com/inspectiontechnologies
Lamp Replacement
CAUTION: Lamp may be Hot! Allow the lamp to cool before replacing.
NOTE: Use only GE Inspection Technologies replacement lamp SPL-600.
AVERTISSEMENT: Pour eviter les rique d’incendie. Utiliser une Lampe de rechange a GE
Inspection Technologies SPL-600.
DO NOT touch the lamp itself or the interior reflective surface, because skin oils will cause
premature lamp failure. If necessary, clean fingerprints with an alcohol dampened cotton
swab. Do not leave lint on the lamp.
1. Turn the unit off and disconnect the power cord from both the wall outlet and the light
source.
2. Loosen finger screw on Lamp Access Panel and pull out Lamp Tray.
3. Remove old lamp by rotating Lamp Spring Retainer away from lamp spring, pulling
lamp back and up and then unplugging lamp connector from harness.
4. Install new lamp by plugging lamp connector into harness and inserting the lamp into
tray housing, ensuring alignment notches are in the 6:00 and 12:00 positions. Rotate
Lamp Spring Retainer back over Lamp Spring. Use only GE Inspection Technologies
replacement lamp SPL-600.
5. Push Lamp Tray back into light box and lock in position with the finger screw.
6. Reconnect power cord.
User Maintenance
Lamp Spring Retainer
Lamp Tray
CAUTION:
Lamp may be Hot!
ELSV-60 Light Source 9