User's Guide

18
Technical specifications
65W AC adaptor
AC input range: 100-240v, 50-60Hz, 2.0A
115V, 400Hz, 1.5A (for aircraft use)
DC output:
24 VDC ± 1 VDC, 2.71A
Typical power consumption: 9.0W
Peak power consumption:
71.9W
Class of Equipment: Class II
Environmental conditions
Operating temperature: +41°F to +95°F (+5°C to +35°C)
Note: The airflow for breathing produced by this therapy device can be
higher than the room temperature. Under extreme ambient temperature
conditions (104°F/40°C) the device remains safe.
Operating humidity: 10 to 95% relative humidity, non-condensing
Operating altitude:
Sea level to 9,870' (3,010 m); air pressure range 700 hPa to 1060 hPa;
Storage pressure/ Storage Altitude 700 to 1060 hPa
Storage and transport temperature:
-13°F to +158°F (-25°C to +70°C)
Storage and transport humidity: 5 to 95% relative humidity, non-condensing
Flow measurement tolerance
±
6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
Pressure measurement tolerance
± 0.5 cm H
2
O ±4% of measured reading
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70 201.12.1.103
Pressure
AirSense 11, humidifier tub
and Standard air tubing
AirSense11, humidifier tub
and SlimLine
AirSense 11, humidifier tub
and ClimateLineAir 11
cm H
2
O (hPa) L/min L/min L/min
4 150 145 144
8
147
142
141
12 143 138 138
16
140
135
134
20 136 131 129
Notes:
Refer to the relevant measurement uncertainty from the table below
Values are after gas standard conversion to STPD (ie, correction factor of 0.9962) and additional 5% guard banding to
buffer the claimed values and rounded down.
Mode pressure range
CPAP 4-20 cm H
2
O (4-20 hPa) (measured at the mask)
CPAP with EPR mode:
4-20 cm H
2
O (4-20 hPa) CPAP with EPR settings: EPR off, Level 1 = 1.0 cm H
2
O,
Level 2 = 2.0 cm H
2
O, Level 3 = 3.0 cm H
2
O
EPR reduces the pressure during expiration by the amount dependent on the
level set above, but the pressure delivered will not drop below 4.0 cm H
2
O.
AutoSet, AutoSet for Her mode: 4-20 cm H
2
O (4-20 hPa)
AutoSet, AutoSet for Her mode with EPR:
4-20 cm H
2
O (4-20 hPa) APAP with EPR settings: EPR off, Level 1 = 1.0 cm H
2
O
Level 2 = 2.0 cm H
2
O,*Level 3 = 3.0 cm H
2
O
EPR reduces the pressure during expiration by the amount dependent on the
level set above, but the pressure delivered will not drop below 4.0 cm H
2
O.