User's Manual

51
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions
IEC 61000-3-2
Class A
Volt age
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
The T10YXXXXXX is
suitable for use in all
establishments, including
domestic establishments
and those directly
connected to the public
low-voltage power supply
network that supplies
buildings used for
domestic purposes.
Guidance and manufacturers declaration –
electromagnetic immunity
The T10YXXXXXX is intended for use in the electromagnetic environm ent specified
below. The customer or the user of the T10YXXXXXX should assure that it is used in
such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
±6 kV
contact
±8 kV
air
±6 kV cont act
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV
for
power
supply
lines
±1 kV
for
input/out
±2 kV for
power
supply lines
±1 kV for
input/output
lines
Mains power
quality should be
that of a typic al
commercial or
hospital
environment.