Owner's manual

1. SAFETY RULES
This instrument is designed for indoor use at temperatures
between 32°F and 122°F (0°C to 50°C), and altitudes up to 6500
ft. (2,000 m), and 10% to 90% relative humidity.
To ensure that the instrument is used safely, follow all safety and
operating instructions in this manual. If the instrument is not used
as described in this operation manual, the safety features of this
device may be impaired.
To avoid personal injury and damage to the instrument use extreme
caution when working around hot or moving objects or machinery
(i.e.: radiators, exhaust manifolds, catalytic converters, drive belts,
y wheels)
2. INTERNATIONAL SYMBOLS
Important information
Refer to the User’s Manual
CAUTION: Laser radiation
AVOID DIRECT EYE EXPOSURE
1mW Output at
630-670 nm
Class II Laser Product
3. TECHNICAL SPECIFICATIONS
3.1 General Specications
Display: 5 digit 0.7(16mm) LCD, max. of 99999
display, with measuring unit indicator
and Laser On Target indicator
Accuracy:
±(0.05%+1 Digit)
Operating temperature:
32°F to 122°F (0°C to 50°C)
Response time: 0.5 sec. (over 120 RPM)
Range selection: Auto Ranging
Memory: Maximum, Minimum, and display
hold.
Distance to target: 2” to 20” (50 mm to 500 mm)
Laser pointer: Laser Diode, < 2mW output
wavelength 630-670 nm, Class II
Power source: 9-volt battery type NEDA 1604,
IEC6F22 IEC 6LR61
Battery life: Approx. 5 hours (w/alkaline batteries)
Power saving: Automatic shut off after 10 seconds
Dimensions: 3.13”x 1.88”x 7.0” (80 x 48 x 178 mm)
Weight: 7.6 oz. (215 g) including battery
Accessories: User’s manual, soft pouch, 3 x 20 cm
reective autoahesive tape and 9 V
alkaline battery
3.2 Measurement Specications
Accuracies are ±(% of reading + number of least signicant dig-
its) at 23°C ±5°C ambient temperature, with less than 75% RH.
4. OPERATION
4.1 Instrument Description
1) Digital display
2) Display backlight button
3) Minimum button
4) Maximum button
5) RPM/Count button
6) Laser output
7) Measument trigger
8) Battery compartment cover
Function Range Acuracy Resolution
RPM
1 to 99,999 RPM ±(0.05%+1 Digit) 1 RPM
COUNT
(total revolutions)
1 to 99,999
±(0.05%+1 Digit)
1 COUNT
General Technologies Corp.
TA110
Laser Optical Tachometer
4.2 Display Description
4.3 Measurement Procedures
!
CAUTION: Avoid contact with moving and/or
hot components, to prevent personal
injury or damage to the instrument.
!
CAUTION: Laser radiation is emitted
Avoid direct eye contact
4.3.1 RPM Measurement
Apply a reective mark or a piece of reective tape to the
rotating object under measurement.
Press the “Measurement trigger” and verify that the “RPM
Indicator” is shown in the display. If the RPM function is not
selected, press the “RPM/COUNT button” once to select the
RPM function.
Press the “Measurement trigger” and aim the laser beam at the
reective mark or tape on the rotating object.
Verify that the “On Target indicator” appears on the display,
indicating that the instrument is detecting the rotation of the
object and the reective mark on it.
Proceed to read the RPM on the digital display.
!
1
2
3
4
5
6
7
8
5
6
7
8
9
10
HOLD indicator
On Target indicator
Depleted battery
warning
MAX indicator
MIN indicator
Backlight indicator
Laser output warning
Numeric reading
RPM indicator
COUNT indicator
Reective
tape
2 to 20”
5 to 50 cm
Laser beam

Summary of content (2 pages)