User guide

5. PH TEMPERATURE COMPENSATION
Enable the meter to gain high accuracy measuring results
from different kinds of solution, the temperature 
compensation calibration procedures are necessary to 
be executed.
Manual temperature compensation calibration procedures
  please refer to 5-1 ( See below).
Automatical temperature compensation calibration procedures
please refer to 5-2 ( Page 15 ).
5-1 Manual temperature compensation procedures
Before the manual temperature compensation calibration 
procedures, please make sure there are no ATC probe 
( YK-200ACT ) in the " Optional Probe Input Socket"
( 3-11, Fig. 1 ).
1) Power on the instrument by pressing the " Power 
Button " ( 3-2, Fig. 1 ).
2) Press the " PH/mV Button " ( 3-6, Fig. 1 ) to select
 the PH function with a " PH " symbol on the display
3) " TEMP. C Button " ( 3-7, Fig. 1 ) is used to adjust the 
following values :
a. Tmperature compensation value 
b. PH 4 Default calibartion value
c. PH 7 Default calibartion value
d. PH 10 Default calibartion value
11