User Manual
HUMIDITY CALIBRATION (8808/09)
Please follow below steps to calibrate
the humidity accuracy of the meter.
1.Turn off the meter, plug the meter probe
 into 33% salt bottle and press 
 "START/STOP" +"SET" at the same
 time over 2sec to enter to calibration 
 mode.
2."32.8%" will flash on the monitor. 
 30 minutes after, the flashing stops to
 indicate the calibration of 33% is 
 finished.
3. Move the meter probe to the 75% salt 
 bottle and press "START/STOP" over 
 2 seconds to enter 75% calibration.
4. "75.3%" will flash on the monitor.
5. lashing stops 
 means the whole calibration has 
 been completed and the calibration 
 data have been saved in memory. 
6. After step5, the meter will return to 
 measurement mode automatically.
30 minutes after, the f
%
%
Note:
a)You can exit calibration without 
 saving by pressing "START/STOP" 
 >2sec before step 5.
b)To get high accuracy, strongly 
 suggest you to do the calibration at 
 23+2 C Temperature environment.
8808
15