User Manual
1.Power on but no display or meter 
 doesn't work .
a)For 8808/06, please check whether 
 the DC power is connected or not.
b)Make sure the time of pressing 
 "START/STOP"  key is more than 
 100 mS.
c)Check the batteries are in place and 
 in good contact and correct 
 polarity.
d)Replace new batteries and try again.
e)Move the batteries for one minute 
 and put them back again
TROUBLESHOOTING
16
2.Calibration failure.
a)For 8808, please check whether 
 the DC power is connected or not.
b)Make sure the calibration bottle is 
 under good sealing status when the 
 meter's probe is plugged into the 
 salt bottle.
c)Make sure the free air temperature is 
 within 23+2 C when calibration is 
 processed.
3. Error Codes.
a)E02: The value is underflow.
b)E03: The value is overflow.
c)E04: The wrong value is caused by 
 E02 or E03.
d)E11: RH calibration error. Need to 
 re-calibrate.
e)E32: IC read/write error. Need to 
 return the meter to dealer for 
 repair.
f)E33: Circuit error in measurement 
 portion. Need to return the meter
 to dealer for repair.